46
MUPs amplitude and duration without a corresponding reference limit are black. The normal area is displayed as a transparent box with a
green border. The dark green bow represents normative values for individual MUPs, the light green box represents normative values for the
mean amplitude/duration.
MUP Analysis (optional)
When clicking the
MUP Analysis button,
the application analyzes the data from the MUP session searching for possible MUPs. Each
unique, repeated
MUP is identified.
The analysis identifies and averages individual discharges from the same motor unit. The action after analysis is predefined and depends on
the Test configuration.
•
Types of actions if MUPs are found:
o
Stay in Current View (auto accept)
o
Go to MUPs
o
Go to MUP Edit
o
Go to MUP Review View
•
Types of actions if no MUPs are found:
o
Stay in the current view (continue)
o
Go to the MUP review view
IP Analysis (optional)
When clicking the
IP analysis button,
the application analyses the data from the IP (Interference Pattern) session, calculates the turns, the
number of small segments, NSS, and other IP parameters.
Motor Unit Potentials (MUP) (optional)
MUP − Data Page
The
Data
page displays a table of the motor unit potentials as shown in the figure below.