RF emissions
CISPR11
Class B
Aurical Otocam 300 is suitable for use in all environments, including domestic environments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
Aurical Otocam 300 is intended for use in the electromagnetic environment specified below. The user of Aurical Otocam 300 should ensure that it is used in such
an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
+/- 1 kV for input/output lines
+/- 1 kV for input/output lines
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems that are NOT life-supporting
Aurical Otocam 300 is intended for use in the electromagnetic environment specified below. The user of Aurical Otocam 300 should ensure that it is used in such
an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Aurical Otocam 300
33
8 Technical Specifications