Electrical data
Supply voltage through USB port
Input power
USB 2.0, Max. 500 mA.
Light source
2 LEDs, fibre-optic light guide
USB plug
USB, type A (LED power supply/Camera/Pre-heating)
Switch activated by magnet contact used to switch between pre-heating and light.
Mechanical data
Cable length
2700 mm (8.86 feet)
Length without cable
170mm ( 6.7 inches)
Weight including cable
250 g (8.8 ounces)
Largest diameter
45 mm (1.8 inches)
Distal diameter
max. 3.4 mm (0.134 inches)
Total weight
1300 g
Service life
Expected Service life
5 years
Storage environment
Temperature
-20°C to +60°C (-4°F to +140°F)
Relative humidity
<90%, non-condensing
Air pressure
500 hPa to 1060 hPa
Operating environment
Temperature
+10°C to +30°C (+50°F to +86°F)
Air humidity
30% to 75%, non-condensing
Air pressure
600 hPa to 1060 hPa
Essential performance
Aurical Otocam 300 has no essential performance.
Standards
Patient Safety
IEC 60601-1:2005+AMD1:2012 and EN 60601-1:2006+A1:2013
CAN/CSA-C22.2 NO. 60601-1:14
Class II; applied part Type BF; IPX0
Endoscopic equipment:
IEC 60601-2-18:2009 and EN 60601-2-18:1996 + A1:2000
30
Aurical Otocam 300
8 Technical Specifications