Appendix A
Specifications
VXI-MXI-Express Series User Manual
A-6
ni.com
Shock and Vibration
Operational shock ...................................30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27; meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating .........................................5 to 500 Hz, 0.3 g
rms
Nonoperating ...................................5 to 500 Hz, 2.4 g
rms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.)
Safety
This product meets the requirements of the following standards of safety
for electrical equipment for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for
electrical equipment for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
In the United States (per FCC 47 CFR), Class A equipment is intended for use in
commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia
and New Zealand (per CISPR 11) Class A equipment is intended for use only in
heavy-industrial locations.