Chapter 1
Installing and Configuring the SCXI-1530/1531
1-12
www.ni.com
module as described in
Removing the SCXI-1530/1531 from
Measurement & Automation Explorer
and add the correct module as
described in
4.
If you get the warning message
Failed to find
, followed by a
module code and then the message
Slot x is empty
, check to see
if the configured module is installed in the specified slot. If not, install
the module by referring to
Installing the SCXI-1530/1531 Module into
. If the module is installed in the correct slot, follow
these steps:
a.
Power off the chassis.
b.
Remove the module as specified in
SCXI-1530/1531 from an SCXI Chassis
c.
Verify that no connector pins are bent on the rear connectors.
d.
Reinstall the module as described in
SCXI-1530/1531 Module into the SCXI Chassis
e.
Ensure the module is fully inserted and properly aligned in
the slot.
5.
Return to
and retest your SCXI chassis.
Removing the SCXI-1530/1531
This section describes how to remove the SCXI-1530/1531 from an SCXI
chassis.
Removing the SCXI-1530/1531 from an SCXI Chassis
Please read
found in this chapter before disconnecting
signals or removing the SCXI-1530/1531 from the chassis.
Note
Only a qualified person who has read and understands all the safety information in
this manual should remove an SCXI module.
You need the following items to complete this task:
•
SCXI chassis or PXI combination chassis with the SCXI-1530/1531
module(s) installed
•
1/4 in. flathead screwdriver
Consult the documentation for your SCXI/PXI chassis and accessories for
additional instructions and warnings. To remove the SCXI-1530/1531