Appendix A
Specifications
SCXI-1502 User Manual
A-2
ni.com
Gain = 100
Calibrated
1
................................±25
µ
V max
±10
µ
V typ
With autozero enabled..............±10
µ
V max
±5
µ
V typ
Gain error (relative to calibration reference)
Gain = 1 or 100
Calibrated
2
................................0.074% of reading max
0.02% of reading typ
Amplifier Characteristics
Input coupling.........................................DC
Input impedance
Normal powered on .........................>1 G
Ω
Input bias current ....................................±
2.8
nA
CMRR characteristics (DC to 60 Hz)
Gain 1 ..............................................–75 dB min
–90 dB typ
Gain 100 ..........................................–106 dB min
–110 dB typ
Output range ...........................................±10 V
Output impedance...................................91
Ω
Dynamic Characteristics
Minimum scan interval (per channel, any gain)
±0.012% accuracy ...........................3
µ
s
±0.0061% accuracy .........................10
µ
s
±0.0015% accuracy .........................20
µ
s
1
25 °C ±10 °C over one year.
2
Assumes 1,000 point average, ±1 °C of autozero temperature.