Chapter 1
About the SCXI-1502
SCXI-1502 User Manual
1-6
ni.com
•
If you get a
Verify SCXI Chassis
message box showing the SCXI
chassis model number,
Chassis ID:
x
, and the message
Slot
Number:
x
Configuration has module: SCXI-
XXXX
or
1502,
hardware in chassis is: SCXI-
YYYY
,
1502
, or
Empty
, complete the
following troubleshooting steps to correct the error.
1.
Expand
NI-DAQmx Devices
.
2.
Right-click the SCXI chassis and click
Properties
to load the
chassis configurator.
3.
Under the
Modules
tab, ensure that the cabled module is listed in
the correct slot.
4.
If the cabled module is not listed in the correct slot, complete the
following troubleshooting steps:
a.
If the cabled module is not listed in the correct slot and the
slot is empty, click the drop-down listbox next to the correct
slot and select the cabled module. Configure the cabled
module following the steps listed in the
SCXI Quick Start
Guide
. Click
OK
.
b.
If another module appears where the cabled module should
be, click the drop-down listbox next to the correct slot and
select the cabled module. A message box appears asking you
to confirm the module replacement. Click
OK
. Configure the
cabled module following the steps listed in the
SCXI Quick
Start Guide
. Click
OK
.
•
If you have more than one kind of SCXI module in the SCXI chassis,
ensure that you have the highest priority SCXI module cabled to the
E/M Series DAQ device. Refer to the
SCXI Quick Start Guide
to find
out which SCXI module in the chassis should be cabled to the
E/M Series DAQ device.
•
After checking the preceding items, return to the
Troubleshooting the
Self-Test Verification
section and retest the SCXI chassis.
If these measures do not successfully configure the SCXI system, contact
NI. Refer to the
Technical Support Information
document for contact
information.