Chapter 2
Configuration and Installation
2-12
www.natinst.com
SCXI-1120 Filter Jumpers
Two-stage filtering is also available on your SCXI-1120 module. The first
stage is located in the isolated section of the input channel, whereas the
second stage is located in the nonisolated section of your input channel.
Two-stage filtering eliminates the noise generated by the isolation
amplifier, producing a higher signal-to-noise ratio. Furthermore, two filter
bandwidths are available, 10 kHz and 4 Hz.
Second-stage
1
A
2
B
5
C
10
D (factory-default setting
for the SCXI-1120)
20
E (factory-default setting
for the SCXI-1120D)
Table 2-6.
SCXI-1120 Filter Jumper Allocation
Input
Channel
Number
First
Filter Jumper
Second
Filter Jumper
4 Hz
(Factory
Default)
10 kHz
4 Hz
(Factory
Default)
10 kHz
0
W17-A
W17-B
W25
W26
1
W18-A
W18-B
W27
W28
2
W19-A
W19-B
W29
W30
3
W20-A
W20-B
W31
W32
4
W21-A
W21-B
W33
W34
5
W22-A
W22-B
W35
W36
6
W23-A
W23-B
W37
W38
7
W24-A
W24-B
W39
W40
Table 2-5.
Gain Jumper Positions (Continued)
Gain
Setting
Jumper Position