10
|
ni.com
|
PXIe-4330/4331 Calibration Procedure
Connect the DMM and Multifunction Calibrator Cables
1.
Connect DMM 1 on the back of the CAL-4330 to a PXI-4071 as shown in Figure 8.
Figure 8.
DMM 1
2.
Connect DMM 2 on the back of the CAL-4330 to another PXI-4071 as shown in Figure 9.
Figure 9.
DMM 2
1
CAL-4330
2
PXI-4071
1
CAL-4330
2
PXI-4071
HI
LO
3
00 V
MAX
CAT II
HI
LO
AUX I/O
1 A
250 V
MAX
3
00 V
MAX
CAT II
5 V
MAX
AMP
S
INPUT
V
˖
S
EN
S
E
˖
4W
NI PXI-4071
6½-Digit FlexDMM
VOLTS
DMM 1
LO
HI
2
1
HI
LO
3
00 V
MAX
CAT II
HI
LO
AUX I/O
1 A
250 V
MAX
3
00 V
MAX
CAT II
5 V
MAX
AMP
S
INPUT
V
˖
S
EN
S
E
˖
4W
NI PXI-4071
6½-Digit FlexDMM
SENSE
DMM 2
LO
HI
VOLTS
LO
HI
1
2