Figure 2. Warranted Current Accuracy Zone for PPMU Force Current
P
ercent of Current Range F
orced (%)
Voltage (V)
-25
75
50
25
0
-50
-75
1 1.25
2
3
-1
0
4
5
100
-2
-100
Zone 1
6
PPMU voltage clamps
Range
-2 V to 6 V
Resolution
122 μV
Accuracy
±100 mV, typical
PPMU Measure Current
Table 2. PPMU Measure Current Accuracy
Range
Resolution
Accuracy
±2 μA
460 pA
±1% of range for Zone 1 of
±1.5% of range for Zone 2 of
±32 μA
7.3 nA
±128 μA
30 nA
±2 mA
460 nA
±32 mA
7.3 μA
8
|
ni.com
|
PXIe-6570 Specifications