Current levels
Range
1.5 mA to 24 mA
Resolution
488 nA
Accuracy
1 mA, 3 V over/under drive, typical
PPMU
PPMU Force Voltage
Signal type
Single-ended, referenced to the DGS pin when
connected. Otherwise referenced to GND.
Voltage levels
Range
-2 V to 6 V
Resolution
122 μV
Accuracy
±15 mV, 1 MΩ load, warranted
PPMU Measure Voltage
Signal type
Single-ended, referenced to the DGS pin when
connected. Otherwise referenced to GND.
Voltage levels
Range
-2 V to 6 V
Resolution
228 μV
Accuracy
±5 mV, warranted
PPMU Force Current
Table 1. PPMU Force Current Accuracy
Range
Resolution
Accuracy
±2 µA
60 pA
±1% of range for Zone 1 of
on page 8, warranted
±32 µA
980 pA
±128 μA
3.9 nA
±2 mA
60 nA
±32 mA
980 nA
PXIe-6570 Specifications
|
© National Instruments
|