background image

Storage Environment

Ambient temperature range

-40 °C to 71 °C (Tested in accordance
with IEC 60068-2-1 and IEC 60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)

Relative humidity range

5% to 95%, noncondensing (Tested in
accordance with IEC 60068-2-56.)

Shock and Vibration

Operating shock

30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)

Random vibration

Operating

5 Hz to 500 Hz, 0.3 g

rms

Nonoperating

5 Hz to 500 Hz, 2.4 g

rms

 (Tested in accordance

with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)

Safety Voltages

Connect only voltages that are within these limits.

Supported measurement range

-2 V to 6 V

Caution

 

Do not connect the system to signals or use for measurements within

Measurement Categories II, III, or IV.

Compliance and Certifications

Safety

This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1, CSA 61010-1

Note

 

For UL and other safety certifications, refer to the product label or the 

Online

Product Certification

 section.

PXIe-6570 Specifications

  | 

© National Instruments

  | 

15

Содержание PXIe-6570

Страница 1: ...ics are unwarranted values that are relevant to the use of the product and convey the expected performance of the product Specifications in this document are Characteristics unless otherwise noted Spe...

Страница 2: ...Voltage 7 PPMU Measure Voltage 7 PPMU Force Current 7 PPMU Measure Current 8 PPMU Programmable Aperture Time 9 Pattern Control 10 Opcodes 10 Pipeline Latencies 11 Source and Capture 12 Frequency Measu...

Страница 3: ...e fly Clock Generation Master clock source PXIe_CLK1003 Clock domains One independent clock domains on a single instrument not supported 1 The system channel count is the maximum number of channels av...

Страница 4: ...te State Non Return NR Pin State Vector Cycle Drive Data Drive Return Drive Data Drive On Drive On Drive Return Drive Data Drive Return Drive Off Drive On Drive Data Drive On Drive Return Drive Data D...

Страница 5: ...Timing Edge placement range Minimum Start of vector period 0 ns Maximum 5 vector periods or 40 s whichever is smaller Minimum required edge separation Between any driven data change 5 ns Between any...

Страница 6: ...tput impedance 50 Rise fall time 20 to 80 1 2 ns up to 5 V Comparator Signal type Single ended referenced to the DGS pin when connected Otherwise referenced to GND Programmable levels VOH VOL Voltage...

Страница 7: ...Accuracy 15 mV 1 M load warranted PPMU Measure Voltage Signal type Single ended referenced to the DGS pin when connected Otherwise referenced to GND Voltage levels Range 2 V to 6 V Resolution 228 V A...

Страница 8: ...PMU voltage clamps Range 2 V to 6 V Resolution 122 V Accuracy 100 mV typical PPMU Measure Current Table 2 PPMU Measure Current Accuracy Range Resolution Accuracy 2 A 460 pA 1 of range for Zone 1 of Fi...

Страница 9: ...Current Percent of Current Range Measured Voltage V 25 75 50 25 0 50 75 1 2 3 1 0 4 5 100 2 100 6 Zone 2 Zone 1 1 25 1 75 PPMU Programmable Aperture Time Aperture time Minimum 4 s Maximum 65 ms Resol...

Страница 10: ...0 100m 190 150 140 10m Pattern Control Opcodes Refer to the following table for supported opcodes Using matched and failed opcode parameters with multiple PXIe 6570 instruments requires the PXIe 6674T...

Страница 11: ...ear_seqflag write_reg Signal set_signal pulse_signal clear_signal Digital Source and Capture capture_start capture capture_stop source_start source Pipeline Latencies Minimum delay between source_star...

Страница 12: ...g equation to calculate the frequency counter error ppm 1 20 1 000 000 where MeasurementTime is the time in seconds over which the frequency counter measurement is configured to run UnknownClockPeriod...

Страница 13: ...with a 1 ms measurement time using the PXIe Clk100 provided by the PXIe 1085 chassis as the timebase Solution 25 1 25 20 1 100 1 000 000 45 Example 2 Calculating Error when Overriding with the PXIe 66...

Страница 14: ...r rails The maximum current drawn from each of these rails can vary depending on the PXIe 6570 mode of operation The total power consumption does not exceed the input power specification Input power 6...

Страница 15: ...60068 2 64 Nonoperating test profile exceeds the requirements of MIL PRF 28800F Class 3 Safety Voltages Connect only voltages that are within these limits Supported measurement range 2 V to 6 V Cauti...

Страница 16: ...spection analysis purposes Note For EMC declarations certifications and additional information refer to the Online Product Certification section CE Compliance This product meets the essential requirem...

Страница 17: ...cycle all NI products must be disposed of according to local laws and regulations For more information about how to recycle NI products in your region visit ni com environment weee RoHS National Inst...

Страница 18: ...legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant...

Отзывы: