background image

Figure 4.  Voltage Measurement Noise for Given Aperture Times, Typical

RMS Noise (V)

10μ

20μ

30μ

0

40μ

50μ

60μ

90μ

100μ

70μ

80μ

110μ

180μ

200μ

120μ

130μ

170μ
160μ

Aperture Time (s)

1m

100μ

10μ

100m

190μ

150μ
140μ

10m

Pattern Control

Opcodes

Refer to the following table for supported opcodes. Using matched and failed opcode
parameters with multiple PXIe-6570 instruments requires the PXIe-6674T synchronization
module. Other uses of flow-control opcodes with multiple PXIe-6570 instruments only require
NI-TCLK synchronization.

10

  | 

ni.com

  | 

PXIe-6570 Specifications

Содержание PXIe-6570

Страница 1: ...ics are unwarranted values that are relevant to the use of the product and convey the expected performance of the product Specifications in this document are Characteristics unless otherwise noted Spe...

Страница 2: ...Voltage 7 PPMU Measure Voltage 7 PPMU Force Current 7 PPMU Measure Current 8 PPMU Programmable Aperture Time 9 Pattern Control 10 Opcodes 10 Pipeline Latencies 11 Source and Capture 12 Frequency Measu...

Страница 3: ...e fly Clock Generation Master clock source PXIe_CLK1003 Clock domains One independent clock domains on a single instrument not supported 1 The system channel count is the maximum number of channels av...

Страница 4: ...te State Non Return NR Pin State Vector Cycle Drive Data Drive Return Drive Data Drive On Drive On Drive Return Drive Data Drive Return Drive Off Drive On Drive Data Drive On Drive Return Drive Data D...

Страница 5: ...Timing Edge placement range Minimum Start of vector period 0 ns Maximum 5 vector periods or 40 s whichever is smaller Minimum required edge separation Between any driven data change 5 ns Between any...

Страница 6: ...tput impedance 50 Rise fall time 20 to 80 1 2 ns up to 5 V Comparator Signal type Single ended referenced to the DGS pin when connected Otherwise referenced to GND Programmable levels VOH VOL Voltage...

Страница 7: ...Accuracy 15 mV 1 M load warranted PPMU Measure Voltage Signal type Single ended referenced to the DGS pin when connected Otherwise referenced to GND Voltage levels Range 2 V to 6 V Resolution 228 V A...

Страница 8: ...PMU voltage clamps Range 2 V to 6 V Resolution 122 V Accuracy 100 mV typical PPMU Measure Current Table 2 PPMU Measure Current Accuracy Range Resolution Accuracy 2 A 460 pA 1 of range for Zone 1 of Fi...

Страница 9: ...Current Percent of Current Range Measured Voltage V 25 75 50 25 0 50 75 1 2 3 1 0 4 5 100 2 100 6 Zone 2 Zone 1 1 25 1 75 PPMU Programmable Aperture Time Aperture time Minimum 4 s Maximum 65 ms Resol...

Страница 10: ...0 100m 190 150 140 10m Pattern Control Opcodes Refer to the following table for supported opcodes Using matched and failed opcode parameters with multiple PXIe 6570 instruments requires the PXIe 6674T...

Страница 11: ...ear_seqflag write_reg Signal set_signal pulse_signal clear_signal Digital Source and Capture capture_start capture capture_stop source_start source Pipeline Latencies Minimum delay between source_star...

Страница 12: ...g equation to calculate the frequency counter error ppm 1 20 1 000 000 where MeasurementTime is the time in seconds over which the frequency counter measurement is configured to run UnknownClockPeriod...

Страница 13: ...with a 1 ms measurement time using the PXIe Clk100 provided by the PXIe 1085 chassis as the timebase Solution 25 1 25 20 1 100 1 000 000 45 Example 2 Calculating Error when Overriding with the PXIe 66...

Страница 14: ...r rails The maximum current drawn from each of these rails can vary depending on the PXIe 6570 mode of operation The total power consumption does not exceed the input power specification Input power 6...

Страница 15: ...60068 2 64 Nonoperating test profile exceeds the requirements of MIL PRF 28800F Class 3 Safety Voltages Connect only voltages that are within these limits Supported measurement range 2 V to 6 V Cauti...

Страница 16: ...spection analysis purposes Note For EMC declarations certifications and additional information refer to the Online Product Certification section CE Compliance This product meets the essential requirem...

Страница 17: ...cycle all NI products must be disposed of according to local laws and regulations For more information about how to recycle NI products in your region visit ni com environment weee RoHS National Inst...

Страница 18: ...legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant...

Отзывы: