Using Microsoft Visual C/C++
If you're using NI-RFmx or NI-RFSA with Microsoft Visual C/C++, locate examples in the
following directories.
Table 13. Location of Microsoft Visual C/C++ Programming Examples
Software Option
Examples Location
NI-RFmx
<NIDocDir>
\RFmx\Demod\Examples
<NIDocDir>
\RFmx\SpecAn\Examples
where
<NIDocDir>
is the
Users\Public\Public Documents
\National Instruments
directory.
NI-RFSA
<NIDocDir>
\NI-RFSA\examples
where
<NIDocDir>
is the
Users\Public\Public Documents
\National Instruments
directory.
Note
LabVIEW examples that demonstrate integration of the NI 5663E with NI
RF vector signal generators and NI toolkit software, including the NI Modulation
Toolkit, are also available online at
.
Troubleshooting
If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit
Installation
Why Is the ACCESS LED Off When the Chassis Is On?
The LEDs may not light until the device has been configured in MAX. Before proceeding,
verify that the NI 5663E appears in MAX.
If the ACCESS LED fails to light after you power on the chassis, a problem may exist with the
chassis power rails, a hardware module, or the LED.
Caution
Apply external signals only while the NI 5663E is powered on. Applying
external signals while the device is powered off may cause damage.
1.
Disconnect any signals from the module front panels.
2.
Power off the chassis.
3.
Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged device.
4.
Reinstall the module in a different chassis slot.
NI PXIe-5663E Getting Started Guide
|
© National Instruments
|
23