manualshive.com logo in svg
background image

Using Microsoft Visual C/C++

If you're using NI-RFmx or NI-RFSA with Microsoft Visual C/C++, locate examples in the
following directories.

Table 13. Location of Microsoft Visual C/C++ Programming Examples

Software Option

Examples Location

 

 

NI-RFmx

<NIDocDir>

\RFmx\Demod\Examples

<NIDocDir>

\RFmx\SpecAn\Examples

where 

<NIDocDir>

 is the 

Users\Public\Public Documents

\National Instruments

 directory.

NI-RFSA

<NIDocDir>

\NI-RFSA\examples

where 

<NIDocDir>

 is the 

Users\Public\Public Documents

\National Instruments

 directory.

Note

 

LabVIEW examples that demonstrate integration of the NI 5663E with NI

RF vector signal generators and NI toolkit software, including the NI Modulation
Toolkit, are also available online at 

ni.com/examples

.

Troubleshooting

If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit 

ni.com/support

.

Installation

Why Is the ACCESS LED Off When the Chassis Is On?

The LEDs may not light until the device has been configured in MAX. Before proceeding,
verify that the NI 5663E appears in MAX.

If the ACCESS LED fails to light after you power on the chassis, a problem may exist with the
chassis power rails, a hardware module, or the LED.

Caution

 

Apply external signals only while the NI 5663E is powered on. Applying

external signals while the device is powered off may cause damage.

1.

Disconnect any signals from the module front panels.

2.

Power off the chassis.

3.

Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged device.

4.

Reinstall the module in a different chassis slot.

NI PXIe-5663E Getting Started Guide

  | 

© National Instruments

  | 

23

Содержание PXIe-5663E

Страница 1: ...5601 NI 5601 RF downconverter module NI PXIe 5622 NI 5622 IF digitizer module NI PXIe 5652 NI 5652 LO source module Caution The protection provided by this product may be impaired if it is used in a...

Страница 2: ...adation install and use this product in strict accordance with the instructions in the product documentation Furthermore any changes or modifications to the product not expressly approved by National...

Страница 3: ...age Caution Never touch the exposed pins of connectors Note Do not install a device if it appears damaged in any way 3 Unpack any other items and documentation from the kit Store the device in the ant...

Страница 4: ...Rigid SMA to SMA Cable Labeled D Part Number 198776A 01 5 Medium Semi Rigid SMA to SMA Cable Labeled C Part Number 198775A 01 6 Flexible SMA to SMA Cable Part Number 190412B 03 7 SMA Driver Bit Part N...

Страница 5: ...ing the NI 5663E in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56 10 to 90 noncondensing Maximum a...

Страница 6: ...ls to test receivers NI RFSA SFP Controls analyzes and presents data similar to stand alone RF vector signal analyzers The NI RFSA SFP operates on the PC so it provides processing storage and display...

Страница 7: ...chassis 3 Inspect the slot pins on the chassis backplane for any bends or damage prior to installation Do not install a module if the backplane is damaged 4 Remove the black plastic covers from all th...

Страница 8: ...ility and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 11 Cover all empty slots using filler panels or fill using slot...

Страница 9: ...er pin of a medium semi rigid SMA to SMA cable labeled C connect the RF OUT connector on the NI 5652 LO source front panel to the LO IN connector on the NI 5601 RF downconverter front panel 2 Using th...

Страница 10: ...F Vector Signal Analyzers Help for more information about how to connect the NI 5663E using an external clock how to connect multiple NI 5663E modules or how to connect the NI 5663E to the NI PXIe 567...

Страница 11: ...nnectors and two LEDs Figure 5 NI 5601 RF Downconverter Module Front Panel 0 dBm NOM RF IN IF OUT LO IN LO OUT 10 MHz 6 6 GHz ACCESS ACTIVE ALL PORTS 5 VDC MAX 50 0 dBm NOM 20 dBm MAX 20 dBm MAX 20 dB...

Страница 12: ...analog RF input signal to be measured by the NI 5663 IF OUT Connects to the IF IN connector on the NI 5622 IF digitizer module front panel This connector is the output terminal for the frequency tran...

Страница 13: ...in order to control the device GREEN The module is ready to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a state other than thermal shutdown or power supply failure...

Страница 14: ...t panel contains four connectors and two LEDs Figure 6 NI 5622 IF Digitizer Module Front Panel ACCESS ACTIVE CLK IN CLK OUT PFI 1 ESD SENSITIVE TTL 6 3 Vp p MAX 2 Vp p NOM 20 dBm MAX 50 50 50 16 Bit I...

Страница 15: ...Module Front Panel Connectors Connector Use IF IN Connects to the IF OUT connector on the NI 5601 RF downconverter module front panel This connector is the input terminal for a frequency translated IF...

Страница 16: ...to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for the Advance Trigger from the configuration list GREEN The module is...

Страница 17: ...0 5 W MAX Reverse Power RF Signal Generator ACCESS ACTIVE Table 8 Device Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or pre...

Страница 18: ...d a problem with a power rail AMBER The module is being accessed Accessed means that the device setup registers are being written to in order to control the device GREEN The module is ready to be prog...

Страница 19: ...identified as not configured in the configuration tree b In the Associated Devices section select the appropriate module from each system component drop down listbox For the NI 5663E you must associat...

Страница 20: ...library files located in one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit NIDir Shared ExternalCompilerSupport C include Windows 10 64 bit 8 1 64 bit 7 64 bit 32 bit installation...

Страница 21: ...s x86 IVI Foundation IVI Microsoft Visual C C Use the header files located in the IVIROOTDIR32 Include directory and import library files located in the IVIROOTDIR32 Lib directory where IVIROOTDIR32 i...

Страница 22: ...an Application with Microsoft Visual C and C topic of the NI RF Vector Signal Analyzers Help if you prefer to manually add all required include and library files to the project Programming Examples Lo...

Страница 23: ...ples Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical support or visit ni com support Installation Why Is the ACCESS LED Off When the Chassis Is...

Страница 24: ...ditions appear in the list If you are using an MXI controller right click PCI to PCI Bridge and select Properties from the shortcut menu to verify that the bridge is enabled What Should I Do if the Mo...

Страница 25: ...nal analyzer 2 Verify the MAX device association for your NI 5663E 3 If you have verified the interconnections and associations of your device and you are still unable to launch the NI RFSA SFP try un...

Страница 26: ...i com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Example Finder Tip The NI RF Vector Signal Analyzers Help is an HTML versio...

Страница 27: ...te for your product at ni com calibration NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 NI also has offices located around the world For telephone suppor...

Страница 28: ...arty legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relev...