manualshive.com logo in svg
background image

Table 1. NI 5663E Software Options

Software Option

Description

Use Case

 

 

 

NI-RFmx

Provides a single-handle
instrument driver with built-in
measurements.

Use NI-RFmx SpecAn to perform
spectral measurements.

Use NI-RFmx Demod to perform
modulation quality measurements,
such as EVM.

Use NI-RFmx cellular personalities
to perform physical layer
measurement analysis on supported
cellular signals.

NI-RFSA
instrument driver

The NI-RFSA APIs each provide
a fixed API for interacting with
your NI 5663E device.

NI-RFSA provides standard IVI-
based functionality needed for
most vector signal analyzer
applications.

Use NI-RFSA to create custom
measurements or applications that
require I/Q data.

Use NI-RFSA along with the
NI Modulation Toolkit to analyze
modulated signals to test receivers.

NI-RFSA SFP

Controls, analyzes, and presents
data similar to stand-alone RF
vector signal analyzers. The
NI-RFSA SFP operates on the
PC, so it provides processing,
storage, and display capabilities.

Use the NI-RFSA SFP to
interactively acquire data.

Installing the Software

You must be an Administrator to install NI software on your computer.
1.

Install an ADE, such as LabVIEW or LabWindows

/CVI

.

2.

Insert the driver software media into your computer. The installer should open
automatically.

If the installation window does not appear, navigate to the drive, double-click it, and
double-click 

autorun.exe

.

3.

Follow the instructions in the installation prompts.

Note

 

Windows users may see access and security messages during

installation. Accept the prompts to complete the installation.

4.

When the installer completes, select 

Restart

 in the dialog box that prompts you to restart,

shut down, or restart later.

6

  | 

ni.com

  | 

NI PXIe-5663E Getting Started Guide

Содержание PXIe-5663E

Страница 1: ...5601 NI 5601 RF downconverter module NI PXIe 5622 NI 5622 IF digitizer module NI PXIe 5652 NI 5652 LO source module Caution The protection provided by this product may be impaired if it is used in a...

Страница 2: ...adation install and use this product in strict accordance with the instructions in the product documentation Furthermore any changes or modifications to the product not expressly approved by National...

Страница 3: ...age Caution Never touch the exposed pins of connectors Note Do not install a device if it appears damaged in any way 3 Unpack any other items and documentation from the kit Store the device in the ant...

Страница 4: ...Rigid SMA to SMA Cable Labeled D Part Number 198776A 01 5 Medium Semi Rigid SMA to SMA Cable Labeled C Part Number 198775A 01 6 Flexible SMA to SMA Cable Part Number 190412B 03 7 SMA Driver Bit Part N...

Страница 5: ...ing the NI 5663E in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56 10 to 90 noncondensing Maximum a...

Страница 6: ...ls to test receivers NI RFSA SFP Controls analyzes and presents data similar to stand alone RF vector signal analyzers The NI RFSA SFP operates on the PC so it provides processing storage and display...

Страница 7: ...chassis 3 Inspect the slot pins on the chassis backplane for any bends or damage prior to installation Do not install a module if the backplane is damaged 4 Remove the black plastic covers from all th...

Страница 8: ...ility and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 11 Cover all empty slots using filler panels or fill using slot...

Страница 9: ...er pin of a medium semi rigid SMA to SMA cable labeled C connect the RF OUT connector on the NI 5652 LO source front panel to the LO IN connector on the NI 5601 RF downconverter front panel 2 Using th...

Страница 10: ...F Vector Signal Analyzers Help for more information about how to connect the NI 5663E using an external clock how to connect multiple NI 5663E modules or how to connect the NI 5663E to the NI PXIe 567...

Страница 11: ...nnectors and two LEDs Figure 5 NI 5601 RF Downconverter Module Front Panel 0 dBm NOM RF IN IF OUT LO IN LO OUT 10 MHz 6 6 GHz ACCESS ACTIVE ALL PORTS 5 VDC MAX 50 0 dBm NOM 20 dBm MAX 20 dBm MAX 20 dB...

Страница 12: ...analog RF input signal to be measured by the NI 5663 IF OUT Connects to the IF IN connector on the NI 5622 IF digitizer module front panel This connector is the output terminal for the frequency tran...

Страница 13: ...in order to control the device GREEN The module is ready to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a state other than thermal shutdown or power supply failure...

Страница 14: ...t panel contains four connectors and two LEDs Figure 6 NI 5622 IF Digitizer Module Front Panel ACCESS ACTIVE CLK IN CLK OUT PFI 1 ESD SENSITIVE TTL 6 3 Vp p MAX 2 Vp p NOM 20 dBm MAX 50 50 50 16 Bit I...

Страница 15: ...Module Front Panel Connectors Connector Use IF IN Connects to the IF OUT connector on the NI 5601 RF downconverter module front panel This connector is the input terminal for a frequency translated IF...

Страница 16: ...to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for the Advance Trigger from the configuration list GREEN The module is...

Страница 17: ...0 5 W MAX Reverse Power RF Signal Generator ACCESS ACTIVE Table 8 Device Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or pre...

Страница 18: ...d a problem with a power rail AMBER The module is being accessed Accessed means that the device setup registers are being written to in order to control the device GREEN The module is ready to be prog...

Страница 19: ...identified as not configured in the configuration tree b In the Associated Devices section select the appropriate module from each system component drop down listbox For the NI 5663E you must associat...

Страница 20: ...library files located in one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit NIDir Shared ExternalCompilerSupport C include Windows 10 64 bit 8 1 64 bit 7 64 bit 32 bit installation...

Страница 21: ...s x86 IVI Foundation IVI Microsoft Visual C C Use the header files located in the IVIROOTDIR32 Include directory and import library files located in the IVIROOTDIR32 Lib directory where IVIROOTDIR32 i...

Страница 22: ...an Application with Microsoft Visual C and C topic of the NI RF Vector Signal Analyzers Help if you prefer to manually add all required include and library files to the project Programming Examples Lo...

Страница 23: ...ples Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical support or visit ni com support Installation Why Is the ACCESS LED Off When the Chassis Is...

Страница 24: ...ditions appear in the list If you are using an MXI controller right click PCI to PCI Bridge and select Properties from the shortcut menu to verify that the bridge is enabled What Should I Do if the Mo...

Страница 25: ...nal analyzer 2 Verify the MAX device association for your NI 5663E 3 If you have verified the interconnections and associations of your device and you are still unable to launch the NI RFSA SFP try un...

Страница 26: ...i com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Example Finder Tip The NI RF Vector Signal Analyzers Help is an HTML versio...

Страница 27: ...te for your product at ni com calibration NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 NI also has offices located around the world For telephone suppor...

Страница 28: ...arty legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relev...

Отзывы: