2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed NI
hardware.
Note
If you are using the NI 5663E with the LabVIEW Real-Time Module,
expand
Remote Systems
. Find your target IP address or name, expand it, and
then expand
Devices and Interfaces
.
Installed modules appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note
If you do not see your module listed, press <F5> to refresh the list of
installed modules. If the module is still not listed, power off the system, ensure
the module is correctly installed, and restart.
4.
Record the identifier MAX assigns to the hardware. Use this identifier when
programming the NI 5663E.
5.
Associate the hardware modules that comprise your device.
a)
Select the NI 5601 that is identified as not configured in the configuration tree.
b)
In the Associated Devices section, select the appropriate module from each system
component drop-down listbox.
For the NI 5663E, you must associate the NI 5622 IF digitizer module and the
NI 5652 LO source module with the NI 5601 RF downconverter.
c)
Click
Save
in the MAX toolbar.
Note
Module associations may be lost when you move the modules to
different chassis slots.
6.
Self-test the device modules by selecting the modules in the configuration tree, and
clicking
Self-Test
in the MAX toolbar. Repeat this step for all modules in your NI 5663E
system.
The MAX self-test performs a basic verification of hardware resources.
Related Information
on page 23
If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit
.
Interconnecting the NI 5663E Modules
on page 8
Refer to the NI RF Vector Signal Analyzers Help for more information about renaming
devices.
NI PXIe-5663E Getting Started Guide
|
© National Instruments
|
19