background image

6

|

ni.com

|

NI PXIe-4339 Calibration Procedure

Test Conditions

The following setup and environmental conditions are required to ensure the NI PXIe-4339 
meets published specifications.

Keep connections to the NI PXIe-4339 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the TB-4339/B/C are secure.

Use shielded copper wire for all cable connections to the TB-4339/B/C. Use twisted-pair 
wire to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ± 5 °C. The NI PXIe-4339 temperature will be 
greater than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 15 minutes to ensure that the NI PXIe-4339 measurement 
circuitry is at a stable operating temperature.

Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters are 
clean, and that the empty slots contain filler panels. For more information, refer to the 

Maintain Forced-Air Cooling Note to Users 

document available at 

ni.com/manuals

.

Initial Setup

Refer to the 

NI PXIe-4339 and TB-4339/B/C Installation Guide and Terminal Block 

Specifications

 for information about how to install the software and hardware and how to 

configure the device in Measurement & Automation Explorer (MAX).

Note

When a device is configured in MAX, it is assigned a device identifier. Each 

function call uses this identifier to determine which DAQ device to verify or to verify 
and adjust. This document uses 

Dev1

 to refer to the device name. In the following 

procedures, use the device name as it appears in MAX.

Verification

The following performance verification procedures describe the sequence of operation and 
provide test points required to verify the NI PXIe-4339. The verification procedures assume that 
adequate traceable uncertainties are available for the calibration references.

The NI PXIe-4339 has eight independent analog input channels. Each channel can be configured 
for voltage mode or ratio mode. Voltage

 

mode is used to measure voltages, and ratio mode is 

used to provide excitation and measure from bridge-based sensors.

You can verify voltage mode for any or all of the channels depending upon your desired test 
coverage. Ratio mode verification requires the use of voltage mode channels to internally 
measure the excitation voltage of the ratio mode channel that is being verified. Voltage mode 
channels used in ratio mode verification require that you determine gain adjust and offset adjust 
values (ratio verification adjustment factors) prior to verifying ratio mode accuracy.

Содержание PXIe-4339

Страница 1: ...nt Procedure 21 EEPROM Update 28 Reverification 29 Specifications 29 Where to Go for Support 29 Software Calibrating the NI PXIe 4339 requires the installation of NI DAQmx on the calibration system Dr...

Страница 2: ...reference information NI PXIe 4339 Specifications NI PXIe 4339 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Infor...

Страница 3: ...select a substitute using the requirements listed in Table 1 Table 1 Recommended Equipment Equipment Recommended Model Requirements Calibrator Fluke 5700 If this instrument is unavailable use a calibr...

Страница 4: ...alog Signal Names Signal Name Signal Description AI Positive input voltage terminal AI Negative input voltage terminal CH2 CH3 CH0 CH1 CH5 CH4 CH6 CH7 PFI0 DGND AOGND AIGND T T T T T T T T T T T T T T...

Страница 5: ...ation Guide and Terminal Block Specifications 2 For voltage verification and adjustment connect the calibrator to the TB 4339 B C as shown in Figure 2 If the calibrator has a guard connection leave it...

Страница 6: ...ow to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Note When a device is configured in MAX it is assigned a device identifier Each function...

Страница 7: ...voltage measurement with the NI PXIe 4339 a Create a DAQmx task b Create and configure the AI channel according to the values shown in Table 3 c Start the task d Average the readings that you acquired...

Страница 8: ...you will acquire a voltage measurement Vmeas for each paired channel at 0 V and 9 V outputs From these results you can determine the ratio verification adjustment factors including the gain adjust Gad...

Страница 9: ...4339 B C as shown in Figure 2 If the calibrator has a guard connection leave it disconnected If the calibrator output is floating you must connect the negative output to AIGND 3 Set the calibrator vo...

Страница 10: ...Ie 4339 a Create a DAQmx task b Create and configure the AI channels according to the values shown in Table 5 c Start the task d Average the readings that you acquired and record the values as chx_9V...

Страница 11: ...ridge V V measurement using the values in Table 6 c Commit the task to turn on and set the excitation voltage using the following parameters Task DAQmx control task task channels in task_out from the...

Страница 12: ...ratiometric reference input value for each ratiometric channel by dividing the calibrator output voltage setting by the paired channels ratio verification adjusted excitation voltage measurement Vm_ex...

Страница 13: ...on V Test Point V V Lower Limit V V Upper Limit V V High Low 0 01 0 01 2 75 0 02475 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 26 9 V V Test Point 1 0 001 26 9 V V 0 Vm_ex_ch calibrator ou...

Страница 14: ...etting V Calibrator Output V Measured Excitation V Test Point V V Lower Limit V V Upper Limit V V High Low 0 02 0 02 2 75 0 0495 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 32 4 V V Test Po...

Страница 15: ...est Point 1 0 001 8 9 V V Test Point 1 0 001 8 9 V V 0 Vm_ex_ch calibrator output Vm_ex_ch 8 9 V V 8 9 V V 0 18 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 8 9 V V Test Point 1 0 001 8 9 V...

Страница 16: ...V Test Point 1 0 001 26 6 V V 0 05 0 05 7 5 0 3375 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 17 7 V V Test Point 1 0 001 17 7 V V 0 Vm_ex_ch calibrator output Vm_ex_ch 17 7 V V 17 7 V V 0...

Страница 17: ...Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 468 8 V V Test Point 1 0 001 468 8 V V 1 1 5 4 5 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 312 2 V V Test Point 1 0 001 312 2 V V 0...

Страница 18: ...calibrator output Vm_ex_ch Test Point 1 0 001 74 V V Test Point 1 0 001 74 V V 0 04 0 04 1 5 0 054 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 49 3 V V Test Point 1 0 001 49 3 V V 0 Vm_ex_...

Страница 19: ...Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 89 0 V V Test Point 1 0 001 89 0 V V 0 08 0 08 1 5 0 108 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 59 3 V V Test Point 1 0 001 59 3...

Страница 20: ...x_ch calibrator output Vm_ex_ch Test Point 1 0 001 133 0 V V Test Point 1 0 001 133 0 V V 0 2 0 2 1 5 0 27 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 88 7 V V Test Point 1 0 001 88 7 V V 0...

Страница 21: ...V V 3 6 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 1561 V V Test Point 1 0 001 1561 V V 4 4 1 5 5 4 Vm_ex_ch calibrator output Vm_ex_ch Test Point 1 0 001 1041 V V Test Point 1 0 001 1041...

Страница 22: ...g you must connect the negative output to AIGND 3 Call the DAQmx Initialize External Calibration function with the following parameters Device In Dev1 Password NI 4 Call the 4339 instance of the DAQmx...

Страница 23: ...nfigured in step 4 9 Repeat steps 4 through 8 for the remaining ranges from Table 16 10 Call the 4339 instance of the DAQmx Adjust SC Express Calibration function with the following parameters calhand...

Страница 24: ...librator output value from step 11 14 Repeat steps 11 through 13 for the remaining calibrator output values from Table 17 15 Repeat steps 10 through 14 for the remaining excitation voltage values from...

Страница 25: ...0 0 18 Table 19 Ratio Mode 0 05 V V Range Range V V Excitation Setting V Calibrator Output V Max Min 0 05 0 05 2 75 0 12375 0 0 12375 0 05 0 05 3 33 0 14985 0 0 14985 0 05 0 05 5 0 225 0 0 225 0 05 0...

Страница 26: ...atio Mode 1 V V Range Range V V Excitation Setting V Calibrator Output V Max Min 1 1 2 75 2 475 0 2 475 1 1 3 33 2 997 0 2 997 1 1 5 4 5 0 4 5 1 1 7 5 6 75 0 6 75 1 1 10 9 0 9 Table 19 Ratio Mode 0 05...

Страница 27: ...ng V Calibrator Output V Max Min 0 04 0 04 0 625 0 0225 0 0 0225 0 04 0 04 1 0 036 0 0 036 0 04 0 04 1 5 0 054 0 0 054 0 04 0 04 2 0 072 0 0 072 0 04 0 04 2 5 0 09 0 0 09 Table 22 Ratio Mode 0 08 V V...

Страница 28: ...5 0 18 0 0 18 Table 23 Ratio Mode 0 2 V V Range Range V V Excitation Setting V Calibrator Output V Max Min 0 2 0 2 0 625 0 1125 0 0 1125 0 2 0 2 1 0 18 0 0 18 0 2 0 2 1 5 0 27 0 0 27 0 2 0 2 2 0 36 0...

Страница 29: ...calhandle output from DAQmx Initialize External Calibration action commit 0 2 0 2 2 5 0 45 0 0 45 Table 24 Ratio Mode 4 V V Range Range V V Excitation Setting V Calibrator Output V Max Min 4 4 0 625...

Страница 30: ...calibration temperature without making any adjustments by initializing an external calibration and closing the external calibration Reverification Repeat the Verification section to determine the As...

Отзывы: