26
|
ni.com
|
NI PXIe-4339 Calibration Procedure
0.05
-0.05
10
0.45
0
-0.45
Table 20.
Ratio Mode (±1 V/V Range)
Range (V/V)
Excitation Setting (V)
Calibrator Output (V)
Max
Min
1
-1
2.75
2.475
0
-2.475
1
-1
3.33
2.997
0
-2.997
1
-1
5
4.5
0
-4.5
1
-1
7.5
6.75
0
-6.75
1
-1
10
9
0
-9
Table 19.
Ratio Mode (±0.05 V/V Range) (Continued)
Range (V/V)
Excitation Setting (V)
Calibrator Output (V)
Max
Min