Notice
Apply external signals only while the PXIe-4162 is powered on.
Applying external signals while the module is powered off may cause
damage.
1. Disconnect any signals from the module front panels.
2. Power off the chassis.
3. Remove the module from the chassis and inspect it for damage. Do not
reinstall a damaged module.
4. Install the module in a different chassis slot from which you removed it.
5. Power on the chassis.
Note
If you are using a PC with a device for PXI remote control
system, power on the chassis before powering on the computer.
6. Verify that the module appears in MAX.
7. Reset the module in MAX and perform a self-test.
Related information
■
What Should I Do if the PXIe-4162 Fails the Self-Test?
1. Restart the system.
2. Launch MAX.
■
Failed self-test
1. Perform self-calibration, then perform the self-test again. The
PXIe-4162 must be calibrated to pass the self-test.
■
Failed self-calibration
1. Perform self-calibration again.
3. Power off the chassis.
© National Instruments
17
PXIe-4162
Содержание PXIe-4162