Status
Indicator
Output Channel State
(Off)
All device outputs are disconnected from their voltage generation sources through
output disconnect relays.
Green
At least one device output is connected to a voltage generation source.
Red
The device has a fault or is in error due to the voltage generated or measured by
the device. Refer to the driver software for possible sources. The device will not
operate until the error is cleared and/or the device is reset.
Table 2. LED Voltage Status Indicator
Pinout
The following figure and table provide detailed signal information for the PXIe-4162
connector.
CH1 Output HI
CH1 Sense HI
CH10 Sense LO
Calibration HI
CH10 Sense HI
CH10 Guard
CH10 Output HI
CH7 Guard
CH7 Sense LO
CH7 Output HI
CH7 Sense HI
CH0 to 11 Output LO
CH4 Guard
CH4 Sense LO
CH4 Output HI
CH4 Sense HI
CH1 Sense LO
CH1 Guard
CH0 Output HI
CH0 Sense HI
CH2 Guard
CH2 Sense LO
CH3 Output HI
CH3 Sense HI
CH5 Guard
CH5 Sense LO
CH6 Output HI
CH6 Sense HI
CH8 Guard
CH8 Sense LO
CH9 Output HI
CH9 Sense HI
CH11 Guard
CH11 Sense LO
CH0 Guard
CH0 Sense LO
NC
CH2 Output HI
CH2 Sense HI
CH3 Guard
CH3 Sense LO
NC
CH5 Output HI
CH5 Sense HI
CH6 Guard
CH6 Sense LO
NC
CH8 Output HI
CH8 Sense HI
CH9 Guard
CH9 Sense LO
NC
CH11 Output HI
CH11 Sense HI
62
61
60
59
56
55
54
57
50
49
48
51
45
44
43
46
20
21
19
18
17
14
13
12
15
8
7
6
9
10
3
2
1
4
42
41
40
39
38
37
36
35
34
33
31
30
29
28
27
26
25
24
23
22
Bundle Channel Pin Signal Description
A
0
62 Output HI
61 Sense HI
© National Instruments
9
PXIe-4162
Содержание PXIe-4162
Страница 1: ...PXIe 4162 2022 07 06...