![National Instruments NI PXIe-5601 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/national-instruments/ni-pxie-5601/ni-pxie-5601_calibration-procedure_3444485019.webp)
NI PXIe-5601 Calibration Procedure
|
© National Instruments
|
19
15. Repeat steps 5 to 14 for all frequencies in Table 10.
16. Close the NI 5601 session using the niRFSA Close VI or the
niRFSA_close
function.
LO Output Power Verification
1.
Connect the RF source to the power splitter input through the 3.5 mm-to-3.5 mm cable.
2.
Connect the power splitter reference output to power sensor A.
3.
Connect a 50
Ω
terminator to the NI 5601 IF OUT connector.
4.
Connect the other power splitter output to the NI 5601 LO IN connector through the
3.5 mm-to-3.5 mm adapter.
Table 10.
RF Frequencies
Start (MHz)
Stop (MHz)
Step (MHz)
10
30
10
40
80
20
300
1,100
100
1,300
1,900
100
2,100
2,300
100
2,500
3,200
100
3,400
3,900
100
4,100
5,700
100
5,900
6,400
100
6,590
—
—
Note
: RF frequencies are selected to avoid false results from spurious responses.
Table 11.
Average Noise Level Verification Test Limits
Frequency
As-Found Limit
As-Left Limit
10 MHz to <30 MHz
<-159 dBm/Hz
<-160 dBm/Hz
30 MHz to <120 MHz
<-159 dBm/Hz
<-160 dBm/Hz
120 MHz to <3 GHz
<-155 dBm/Hz
<-156 dBm/Hz
3 GHz to <5 GHz
<-153 dBm/Hz
<-154 dBm/Hz
5 GHz to 6.6 GHz
<-151 dBm/Hz
<-152 dBm/Hz