NI PXIe-5601 Calibration Procedure
|
© National Instruments
|
25
3.
Connect the power splitter reference output to power sensor A.
4.
Ensure that the 50
Ω
terminator is connected to the NI 5601 LO OUT connector.
5.
Connect the other power splitter output to the NI 5601 RF IN connector through the
3.5 mm-to-3.5 mm adapter. Figure 8 shows the completed equipment setup.
Figure 8.
Gain Compression Equipment Setup
Note
The figure shows NI PXI-5652, but you can also use NI PXIe-5652.
6.
Create a new session for the NI 5601 using the niRFSA Initialize With Options VI or the
niRFSA_initWithOptions
function. Set the Driver Setup portion of the
option string
parameter to
DriverSetup=Digitizer:<external>
.
1