NI PXIe-5601 Calibration Procedure
5.
Set the NI 5601 reference level according to Table 8 using the niRFSA Configure
Reference Level VI or the
niRFSA_ConfigureReferenceLevel
function.
6.
Set the RF attenuation according to Table 8 using the niRFSA Attenuation property or the
NIRFSA_ATTR_ATTENUATION
attribute.
7.
Set the NI 5601 center frequency according to Table 9 and set the span according to Table 8
using the Center Span instance of the niRFSA Configure Spectrum Frequency VI or the
niRFSA_ConfigureSpectrumFrequencyCenterSpan
function.
8.
Commit the settings to hardware using the niRFSA Initiate VI or the
niRFSA_Initiate
function.
9.
Set the spectrum analyzer using the following settings:
•
Reference level: -50 dBm
•
Frequency span: 100 Hz
•
Resolution bandwidth: 3 kHz
•
Video bandwidth: 10 Hz
•
Center frequency: Refer to Table 9
10. Use the spectrum analyzer noise marker to measure the average noise.
11. Record the average noise measured on the spectrum analyzer. This value is the
Measured
Noise
and is used in step 13.
12. Recall the
NI 5601 Stored Gain
for each setting using the niRFSA Downconverter Gain
property or the
NIRFSA_ATTR_DOWNCONVERTER_GAIN
attribute.
13. Calculate the
NI 5601 Input Noise
using the following formula:
NI 5601 Input Noise
=
Measured Noise
+
IF Calibration Factor
-
NI 5601 Stored Gain
14. Compare the
NI 5601 Input Noise
to the limits in Table 11.
Table 8.
NI 5601 Settings
Frequency Range
Reference
Level
RF Attenuation
Span
10 MHz to 30 MHz
-10 dBm
0 dB
10 Hz
30 MHz to 6,600 MHz
-50 dBm
0 dB
10 Hz
Table 9.
Center Frequency Settings
NI 5601 Center Frequency
Spectrum Analyzer
Center Frequency
10 MHz to <120 MHz
187.5 MHz
120 MHz to <330 MHz
53 MHz
330 MHz to 6.6 GHz
187.5 MHz