Appendix A
Specifications
©
National Instruments Corporation
A-3
NI PXI-6682 Series User Manual
Synchronized trigger clock
to out time, t
co
......................................... 10 ns, typical (relative to
CLKOUT when set up to route
PXI_CLK10)
Output current ........................................ ±48 mA, max
Square wave rise/fall time
(10 to 90%) for 50
Ω
load..................... 0.5 ns min,
2.5 ns max
Input Characteristics
1
Frequency range ..................................... DC to 30 MHz
Input impedance ..................................... 1 k
Ω
, nominal
Input coupling ........................................ DC
Voltage level .......................................... 0 to +3.3 V, +5 V tolerant
Absolute maximum input voltage
2
......... –0.5 V to 6.0 V
Input threshold
Voltage threshold high.................... +2.3 V
Voltage threshold low ..................... +0.8 V
Asynchronous delay, t
pd
PFI<0..2> to
PXI_TRIG<0..7> output ................. 20 to 31 ns, typical
PFI<0..2> to
PXI_STAR<0..12> output
3
............. 10 ns, typical
Synchronized trigger
input setup time, t
setup
............................. 19 ns, typical (relative to
CLKOUT when set up to route
PXI_CLK10)
1
For PFI 0 these characteristics apply when the line is configured as a digital input. They do not apply when configured as an
IRIG-B AM input.
2
Stresses beyond those listed can cause permanent damage to the device. Exposure to absolute maximum rated conditions for
extended periods of time can affect device reliability. Functional operation of the device outside the conditions indicated in
the operational parts of the specifications is not implied.
3
The NI PXI-6682H does not have star trigger lines.