For the best system accuracy, observe the following conditions
when connecting to the NI 9236.
•
Set up the connections to EXC and RC with equal lengths of
an identical wire type and gauge.
•
Connect the AI terminal directly at the sensor instead of
shorting AI to RC directly at the terminals.
High-Vibration Application Connections
If your application is subject to high vibration, NI recommends
that you use a backshell kit to protect connections to the NI 9236.
Overvoltage Protection
The NI 9236 provides overvoltage protection between any two
terminals.
Note
Refer to the device datasheet on
ni.com/manuals
for more information about overvoltage protection.
12
|
ni.com
|
NI 9236 Getting Started Guide