built into the product if the product is damaged in any
way. If the product is damaged, return it to NI for
repair.
Safety Voltages
Connect only voltages that are within the following limits:
Between any two terminals
±30 V maximum
Isolation
Channel-to-channel
None
Channel-to-earth ground
Continuous
60 V DC,
Measurement Catergory I
Withstand
1,000 V RMS, verified by a 5 s
dielectric withstand test
Measurement Category I is for measurements performed on
circuits not directly connected to the electrical distribution system
referred to as
MAINS
voltage. MAINS is a hazardous live
electrical supply system that powers equipment. This category is
for measurements of voltages from specially protected secondary
circuits. Such voltage measurements include signal levels, special
NI 9236 Getting Started Guide
|
© National Instruments
|
3