Test Equipment
The following table lists the equipment recommended for the performance verification and
adjustment procedures. If the recommended equipment is not available, select a substitute
using the requirements listed.
Table 1.
Recommended Equipment
Equipment
Recommended Model
Requirements
Calibrator
Fluke 5502A, Fluke 5522A,
Fluke 5730A
Use a high-precision voltage
source with gain error of
<100 ppm and offset error
<0.1 mV.
BNC (F) to Banana
Adapter
Pomona Electronics 1270
—
2
|
ni.com
|
NI 9231 Calibration Procedure