Chapter 3
Device Overview and Theory of Operation
3-12
ni.com
In below-low-level triggering mode, shown in Figure 3-8, the trigger is
generated when the signal value is less than
lowValue
.
highValue
is
unused.
Figure 3-8.
Below-Low-Level Triggering Mode
In above-high-level triggering mode, shown in Figure 3-9, the trigger is
generated when the signal value is greater than
highValue
.
lowValue
is
unused.
Figure 3-9.
Above-High-Level Triggering Mode
In inside-region triggering mode, shown in Figure 3-10, the trigger is
generated when the signal value is between
lowValue
and
highValue
.
Figure 3-10.
Inside-Region Triggering Mode
lowValue
Trigger
highValue
Trigger
highValue
Trigger
lowValue