Chapter 1
Getting Started with the NI 447X
1-6
ni.com
sub-bus are disabled by default and not ever enabled. Damage may result if
these lines are driven by the sub-bus.
In some circumstances, it is possible to drive some of these lines in the
same chassis as the NI PXI-4472. For more information, search the
NI Developer Zone at
ni.com/zone
.
Safety Information
The following section contains important safety information that you
must
follow when installing and using NI 447
X
.
Do
not
operate the NI 447
X
in a manner not specified in this document.
Misuse of the NI 447
X
can result in a hazard. You can compromise the
safety protection built into the NI 447
X
if the NI 447
X
is damaged in any
way. If the NI 447
X
is damaged, return it to NI for repair.
Do
not
substitute parts or modify the NI 447
X
except as described in this
document. Use the NI 447
X
only with the chassis, modules, accessories,
and cables specified in the installation instructions. You
must
have all
covers and filler panels installed during operation of the NI 447
X
.
Do
not
operate the NI 447
X
in an explosive atmosphere or where there may
be flammable gases or fumes. If you must operate the NI 447
X
device in
such an environment, it must be in a suitably rated enclosure.
If you need to clean the NI 447
X
, use a soft, nonmetallic brush. Make sure
that the NI 447
X
is completely dry and free from contaminants before
returning it to service.
Operate the NI 447
X
only at or below Pollution Degree 2. Pollution is
foreign matter in a solid, liquid, or gaseous state that can reduce dielectric
Table 1-1.
J2 Connector Pins Used by the NI PXI-4472
NI PXI-4472 Signal
PXI Pin Name
PXI J2 Pin Number
Master Clock Distribution
LBL<0..12>
C20, E20, A19, C19, D19, E19, D15,
D2, E2, A1, C1, D1, E1
Sync Pulse
PXI Trigger 5
C18
Reserved
LBR<0..12>
A21, C21, D21, E21, A20, B20, E15,
A3, C3, D3, E3, A2, B2
Master Clock Reception
PXI Star In
D17