SCXI-1125 Calibration Procedure
8
ni.com
5.
Add a voltage channel using the DAQmx Create Virtual Channel VI.
6.
Set the SCXI-1125 to a gain setting that corresponds to the current test
point in Table 2 using the DAQmx Channel Property Node. You can
find the AI.Gain property by selecting
Analog Input»General
Properties»Advanced»Gain
.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCreateAIVoltageChan
with the
following parameters:
taskHandle
:
taskHandle
physicalChannel
:
"SC1Mod1/ai
X
"
†
nameToAssignToChannel
:
"
myVoltageChannel
"
terminalConfig
:
DAQmx_Val_Cfg_Default
minVal
: -5 V
maxVal
: 5 V
units
:
DAQmx_Val_Volts
customScaleName
:
""
†
Where
X
refers to the channel you are verifying.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetAIGain
with the following
parameters:
taskHandle
:
taskHandle
channel
:
""
data
:
(gain value for particular test
point) float64