
SCXI-1125 Calibration Procedure
4
ni.com
Calibration Process
The calibration process has the following steps:
1.
—Configure the SCXI-1125 for calibration.
2.
—Verify the existing operation of the
SCXI-1125. This step determines whether the SCXI-1125 is operating
within its test limits.
3.
—Perform an external calibration that adjusts
the SCXI-1125 calibration constants with respect to a known voltage
source.
4.
—Perform another verification to ensure
that the SCXI-1125 is operating within its test limits after adjustments.
Initial Setup
Complete the following steps to configure the SCXI-1125 for calibration:
1.
Make sure that all the appropriate driver and application software is
installed.
2.
Make sure all components involved in the calibration procedure are
powered off.