SCXI-1125 Calibration Procedure
14
ni.com
50
-0.1
-0.099000
-0.098331
-0.099669
100
0.05
0.0495000
0.049846
0.049154
100
0
0.000000
0.000091
-0.000091
100
-0.05
-0.049500
-0.049154
-0.049846
200
0.025
0.024750
0.024936
0.024564
200
0
0.000000
0.000058
-0.000058
200
-0.025
-0.024750
-0.024564
-0.024936
250
0.02
0.019800
0.019954
0.019646
250
0
0.000000
0.000051
-0.000051
250
-0.02
-0.019800
-0.019646
-0.019954
500
0.01
0.009900
0.009989
0.009811
500
0
0.000000
0.000038
-0.000038
500
-0.01
-0.009900
-0.009811
-0.009989
1000
0.005
0.00495
0.005007
0.004893
1000
0
0.000000
0.000031
-0.000031
1000
-0.005
-0.004950
-0.004893
-0.005007
2000
0.0025
0.002475
0.002515
0.002435
2000
0
0.000000
0.000027
-0.000027
2000
-0.0025
-0.002475
-0.002435
-0.002515
Limits valid for 4 Hz or 10 kHz filter settings.
Valid when using an MIO device after performing an MIO internal calibration.
Valid for both filter settings with 1,000 point averaging at 1,000 samples per second.
Table 2.
Upper and Lower Limit Channel Settings (Continued)
Gain
Range
Test Point
Upper Limit (V)
Lower Limit (V)