Configure TestStand to include the following search paths, including subdirectories, in
the search directories:
–
C:\Program Files\National Instruments\LabVIEW 2018\user.lib
\OTA Test
–
C:\Program Files\National Instruments\LabVIEW 2018\instr.lib
–
C:\Program Files\National Instruments\LabVIEW 2018\vi.lib
\RFmx
•
Configure TestStand to call LabVIEW Development System (64-bit).
•
Autonomous mode uses PXI trigger lines between the PXIe-6361 and PXIe-5831. Route
the following trigger routes in MAX:
Note
The PXIe-6361 appears as
DAQ
in MAX.
–
PXI_Trig0 from the PXIe-6361 bus segment to the PXIe-5831 bus segment
–
PXI_Trig1 from the PXIe-5831 bus segment to the PXIe-6361 bus segment
Complete the following steps to perform a full antenna pattern test in autonomous mode.
1.
Navigate to
C:\Users\Public\Public Documents\National Instruments
\TestStand 2019 (64-bit)\Examples\OTA Test\Sequences
and open the
CSV settings file,
Autonomous_TRP.csv
.
2.
Verify that the individual modules are named correctly in MAX. If the names do not
match, either change the name in MAX to match the file or change the file to match your
MAX settings.
3.
Open the
Autonomous_TRP.seq
TestStand sequence, located at
C:\Users\Public
\Public Documents\National Instruments\TestStand 2019 (64-bit)
\Examples\OTA Test\Sequences
.
4.
Run the sequence.
Notice
Verify that nothing interferes with the motion of the positioner. The
sequence triggers motion which may cause damage if anything interferes with
the positioner.
The sequence runs for about 30 seconds collecting data at 2° intervals elevation and 5°
intervals azimuth.
Use the mmWave OTA Test Visualizer to visualize the data in the CSV file that is created in
the Results folder, located in the same directory as the Sequences folder.