M2.5x8 screws (x2)
–
M2.5x6 screws (x2)
•
AIO-3691 High-Speed Serial Adapter Module
2
, with the following components:
–
Female-to-female standoffs (x2)
–
M2.5x8 screws (x2)
–
M2.5x6 screws (x4)
Preparing the Environment
Ensure that the environment you are using the ATCA-3671 in meets the following
specifications.
Ambient temperature range
(IEC 60068-2-1, IEC 60068-2-2)
0 °C to 40 °C
Operating temperature range
Used with a Single-Module ATCA
Chassis
0 °C to 25 °C
Used with a 14-Slot ATCA Chassis
Dependent on final system installation
Note
Operating temperatures are valid only when the ATCA-3671 module is used
with the specified chassis.
Operating humidity (IEC 60068-2-56)
10% to 90% RH, noncondensing
Pollution Degree
2
Maximum altitude
2,000 m at 25 °C ambient temperature
Indoor use only.
Caution
Electrostatic Discharge (ESD) can damage the AIO-3681/3682 and
AIO-3691. To prevent damage, use industry-standard ESD prevention measures
during installation, maintenance, and operation.
Installing the Software
You must be an Administrator to install NI software on your computer.
1.
Install an ADE, such as LabVIEW.
2.
If you chose to install LabVIEW in step 1, install LabVIEW FPGA.
3.
Visit
atcadriver
to access the driver download
page for the latest NI ATCA FPGA Modules software.
4.
Download the NI ATCA FPGA Modules driver software.
2
Included in some ATCA-3671 kits.
ATCA-3671 Getting Started Guide
|
© National Instruments
|
5