© National Instruments
|
7-41
Using an External Source
You can route any of the following signals as Counter
n
Sample Clock:
•
PFI <0..15>
•
RTSI <0..7>
•
PXI_STAR
•
PXIe_DSTAR<A,B>
•
Analog Comparison Event
You can sample data on the rising or falling edge of Counter
n
Sample Clock.
Routing Counter
n
Sample Clock to an Output Terminal
You can route Counter
n
Sample Clock out to any PFI <0..15> terminal. The PFI circuitry inverts
the polarity of Counter
n
Sample Clock before driving the PFI terminal.
Counter
n
Internal Output and Counter
n
TC Signals
The Counter
n
Internal Output signal changes in response to Counter
n
TC.
The two software-selectable output options are pulse output on TC and toggle output on TC. The
output polarity is software-selectable for both options.
With pulse or pulse train generation tasks, the counter drives the pulse(s) on the Counter
n
Internal Output signal. The Counter
n
Internal Output signal can be internally routed to be a
counter/timer input or an “external” source for AI, AO, DI, or DO timing signals.
Routing Counter
n
Internal Output to an Output Terminal
You can route Counter
n
Internal Output to any PFI <0..15>, RTSI <0..7>, or PXIe_DSTARC
terminal. All PFIs are set to high-impedance at startup.
Frequency Output Signal
The Frequency Output (FREQ OUT) signal is the output of the frequency output generator.
Routing Frequency Output to a Terminal
You can route Frequency Output to any PFI <0..15> or PXIe_DSTARC terminal. All PFIs are
set to high-impedance at startup. The FREQ OUT signal can also be routed to DO Sample Clock
and DI Sample Clock.
Содержание 6368