7-38
|
ni.com
Chapter 7
Counters
Routing a Signal to Counter
n
Gate
Each counter has independent input selectors for the Counter
n
Gate signal. Any of the following
signals can be routed to the Counter
n
Gate input:
•
RTSI <0..7>
•
PFI <0..15>
•
AI Reference Trigger (ai/ReferenceTrigger)
•
AI Start Trigger (ai/StartTrigger)
•
AO Sample Clock (ao/SampleClock)
•
DI Sample Clock (di/SampleClock)
•
DI Reference Trigger (di/ReferenceTrigger)
•
DO Sample Clock (do/SampleClock)
•
PXI_STAR
•
PXIe_DSTAR<A,B>
•
Change Detection Event
•
Analog Comparison Event
In addition, a counter’s Internal Output or Source can be routed to a different counter’s gate.
Some of these options may not be available in some driver software.
Routing Counter
n
Gate to an Output Terminal
You can route Counter
n
Gate out to any PFI <0..15>, RTSI <0..7>, or PXIe_DSTARC terminal.
All PFIs are set to high-impedance at startup.
Counter
n
Aux Signal
The Counter
n
Aux signal indicates the first edge in a two-signal edge-separation measurement.
Routing a Signal to Counter
n
Aux
Each counter has independent input selectors for the Counter
n
Aux signal. Any of the following
signals can be routed to the Counter
n
Aux input:
•
RTSI <0..7>
•
PFI <0..15>
•
AI Reference Trigger (ai/ReferenceTrigger)
•
AI Start Trigger (ai/StartTrigger)
•
PXI_STAR
•
PXIe_DSTAR<A,B>
•
Analog Comparison Event
•
Change Detection Event
Содержание 6368
Страница 1: ...PXIe 6368...