NI 6110/6111/6115/6120 Calibration Procedure
18
ni.com
±1
Bipolar
10
NegFS
–0.99000
–0.99120
–0.98880
–0.99122
–0.98878
±0.5
Bipolar
20
PosFS
0.495000
0.494302
0.495698
0.494292
0.495708
±0.5
Bipolar
20
Zero
0.000000
–0.000411
0.000411
–0.000411
0.000411
±0.5
Bipolar
20
NegFS
–0.495000
–0.495698
–0.494302
–0.495704
–0.494292
±0.2
Bipolar
50
PosFS
0.198000
0.197607
0.198393
0.197603
0.198397
±0.2
Bipolar
50
Zero
0.000000
–0.000189
0.000189
–0.000189
0.000189
±0.2
Bipolar
50
NegFS
–0.198000
–0.198393
–0.197607
–0.198397
–0.197603
Table 9.
Analog Output Values for the NI 6115
Range
Polarity
Test Point
24-Hour Ranges
1-Year Ranges
Location
Value (V)
Lower
Limit (V)
Upper
Limit (V)
Lower
Limit (V)
Upper
Limit (V)
20
Bipolar
PosFS
9.990000
9.976734
10.003266
9.976565
10.003435
20
Bipolar
Zero
0.000000
–0.008900
0.008900
–0.008900
0.008900
20
Bipolar
NegFS
–9.990000
–10.003266
–9.976734
–10.003435
–9.976565
Table 10.
Counter Values for the NI 6115
Counter Values
Set Point (MHz)
Lower Limit (MHz)
Upper Limit (MHz)
—
5
5.0005
4.9995
Table 11.
Analog Input Values for the NI 6120
Range
Polarity
Gain
Test Point
24-Hour Ranges
1-Year Ranges
Location
Value (V)
Lower
Limit (V)
Upper
Limit (V)
Lower
Limit (V)
Upper
Limit (V)
±42
Bipolar
0.20
PosFS
41.5800
41.506321
41.653679
41.505621
41.654379
±42
Bipolar
0.20
Zero
0.00000
–0.008303
0.008303
–0.008303
0.008303
±42
Bipolar
0.20
NegFS
–41.5800
–41.653679
–41.506321
–41.654379
–41.505621
±20
Bipolar
0.50
PosFS
19.8000000
19.769172
19.830828
19.768839
19.831161
±20
Bipolar
0.50
Zero
0.00000000
–0.003322
0.003322
–0.003322
0.003322
Table 8.
Analog Input Values for the NI 6115 (Continued)
Range
Polarity
Gain
Test Point
24-Hour Ranges
1-Year Ranges
Location
Value (V)
Lower
Limit (V)
Upper
Limit (V)
Lower
Limit (V)
Upper
Limit (V)