NI 4065 Calibration Procedure
You have completed verifying AC voltage for the NI 4065. Select one of the
following options:
•
If you want to continue verifying other modes, go to the
section.
•
If you do
not
want to verify any additional modes
and
you are
performing a
pre-adjustment
verification, call the niDMM Close VI or
the
niDMM_close
function to close the session.
8
2
2
20 kHz
9
2
2
50 kHz
10
2
2
100 kHz
11
20
20
10 Hz
12
20
20
40 Hz
13
20
20
20 kHz
14
20
20
50 kHz
15
20
20
100 kHz
16
300
200
10 Hz
17
300
200
40 Hz
18
300
200
20 kHz
19
300
200
50 kHz
20
300
200
100 kHz
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_close
with the following
parameter:
Instrument_Handle
: The instrument
handle from
niDMM_init
Table 6.
AC Voltage Flatness Settings (Continued)
Iteration
Range (V
ac
)
Calibrator Output
Amplitude (V
ac
)
Frequency