NI 4065 Calibration Procedure
36
ni.com
Verifying 2-Wire Resistance
To verify 2-wire resistance for the NI 4065, complete the following steps:
1.
Reset the calibrator.
2.
Connect the NI 4065 and the Fluke 5700A/5720A calibrator using
two sets of Fluke 5440 cables, a double banana plug, and
two insulated, low EMF spade lugs, as shown in Figure 5 for the
NI PXI/PCI/PCIe-4065 and Figure 6 for the NI USB-4065. Table 9
lists the cable connections.
Figure 5.
NI PXI/PCI/PCIe-4065 Cable Connections for 2-Wire Resistance
1
NI PCI-4065
2
Fluke 5440 Cables
5720A HI / LO
S
EN
S
E
5720A HI / LO
CAT II
INPUT
300V
HI
LO
AMPS
3A
SENSE
300V
AUX
I/O
5V
HI
LO
2
1