NI 4065 Calibration Procedure
30
ni.com
Verifying 4-Wire Resistance
To verify the 4-wire resistance of the NI 4065, complete the following
steps:
1.
Reset the calibrator.
2.
Connect the NI 4065 and the Fluke 5700A/5720A calibrator using the
Fluke 5440 cables, as shown in Figure 3 for the
NI PXI/PCI/PCIe-4065 and Figure 4 for the NI USB-4065. Table 7
lists the cable connections.
Figure 3.
NI PXI/PCI/PCIe-4065 Cable Connections for 4-Wire Resistance
1
NI PCI-4065
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cables