Male Catheterization
& Prostate Examination
Features included with
LF04301, LF04302, and
LF04302EX .
The male genital
insert can represent
an uncircumcised or
circumcised adult male,
with the addition of
the included foreskin . To change genitals in the
simulator, simply press to fit into the open genital
area of the pelvis . (See Figure 29) .
The urinary catheterization reservoir is
interchangeable between male and female genitals .
Female Genitals come installed with the simulator .
Note: To apply the foreskin, apply the included pump
spray lubricant to the glans and slide the foreskin in
place .To prepare for catheterization exercises:
1 . The urinary resevoir is installed on the back side,
make sure the urinary
resevoir is firmly in
place . (See Figure 30) .
2 . Close the clamp on
the Fluid Supply Bag
(See Figure 31) and
fill with water or
included simulated
urine . Recommended
amount of fluid to fill
the Fluid Supply Bag
is between 100 and
250 mL .
3 . Hang the filled Fluid
Supply Bag . Attach the
tubing quick connect
on end of Fluid Supply
Bag to the quick
connect on the left
lower side of the simulator’s torso .Open Clamp
on fluid supply bag (See Figure 32) .
Set Up
Male Catheterization
Fig 33.
Fig 32.
Fig 29.
Note: The Fluid Supply
Stand is not included .
Use distilled water or
follow instructions on
the Simulated Urine .
4 . Lubricate the
catheter being used
with the Pump Spray
Lubricant provided
(See Figure 33) . The
catheter should be
well lubricated and
force should not be
used when placing
the tube .
Note: Special care
should be taken when
using a Foley catheter .
Nasco recommends
use of 16 French Foley
catheters . Use of this size will avoid the possibility
of leakage . Cuff inflation should only be attempted
when the catheter is in the proper position inside
the bladder . The cuff must also be completely
deflated before the catheter is removed . The catheter
should not be left inserted in the simulator for an
extended period of time . Improper use of a Foley
catheter may result in damage to the simulator and
void the warranty .
5 . After completion of the exercise, close the clamp
on the Fluid Supply Bag tubing and remove the
male genital insert by reversing the assembly
instructions . Disconnect the urinary catheter
reservoir tubing via the quick disconnect, remove
the reservoir from the male genitalia insert, and
drain the reservoir thoroughly (See Figure 30) .
6 . Rinse the reservoir and the outside of the male
genitalia to remove any residual lubricant .
Fluids should be drained and the reservoir emptied
after every use . Ensure the reservoir is clean and dry
prior to storing .