Operating PMM Networks
7-1
7 - Operating PMM Networks
7.1 Operating PMM
Networks
To operate the PMM Artificial Mains Networks connect the IEC plug on the
rear input to the mains insulating transformer.
To avoid any damage caused by transient current pulses, connect the RF
cable to the receiver only after the AC supply has been switched on.
The Equipment Under Test should be connected to the D.U.T. connector.
To select the phase line to be tested operate the phase knob on the front
panel, in this case there is no led indication.
Using PMM receivers the phase switching can be automatically performed;
in this case the phase under test will be indicated by the led in the front
panel.
To avoid any damage caused by transient current pulses, disconnect
the test Signal Analyzer input before switching AC supply on or off
To allow correct equipment ventilation ensure that all the vent grids
on the equipment covers be free by any obstructing object.
The PMM LISN continuous operation at higher current limit shall not
exceed one hour, a rest period of at least a half hour shall be
observed to allow the equipment to cool down.