Calibrating
1. Press
Settings
and then press
Calibration button
.
2. Check the current background level and press
Yes button
to
recalibrate.
3. To recalibrate the EVE™, mix 10 µL trypan blue solution with
10 µL of a standard buffer, such as phosphate buffered saline
(PBS). Mix thoroughly.