Page 18
17-110 -0143
Press the Enter button ( )
Press the Menu button ( )
Display should read E:1
Press the Menu button ( ) to close the
Menu system. The sensor will revert back to
normal monitoring mode.
Using the Next (increment) button ( )
and the Previous (decrement) button ( ),
set the display to that of the calibration
gas level.
Press the Enter button ( ). This will
“Span” the Sensor.
The display will read (- - - -) to confirm
that the Sensor Span been successful
7.25 psi
(0.5 bar)
250ml/m
9.3 Sensor Calibration Procedure