English
User's Manual
of GNF001
Version 1, July 2023
- 4 -
1
th
Terms of Use Security
The company. low-speed BenchtopAir-refrigerate Centrifuge is based on current
technology and safety standards:
GB4793.1-2007 Measurement, control electrical equipment for laboratories safety
requirements part 1: General safety requirements
;
GB4793.7-2001 Measurement, control electrical equipment for laboratories safety
requirement special requirements for laboratory centrifuge
;
GB191-2000 Transportation packaging logo icon
;
GB6587.6-86 Transport-Test of Electronic Measurement Instruments
;
GB/T14710-1993 Medical electrical equipment environmental requirements and
test methods
;
Therefore, using must be in accordance with the requirements of the design. If
incorrect or inappropriate use happens as follows when using The company.
low-speed BenchtopAir-refrigerate Centrifuge, it will result in equipment damage or
personal injury
:
●
Use not in accordance with design requirements;
●
Operation and maintenance staff to use without training;
●
User changes content inappropriately without authorization of the design
;
●
Not pay attention to or understand the rules of safe use.
Any person who involving the using or maintenance of
The company. low-speed BenchtopAir-refrigerate
Centrifuge should read and understand the using way and
rules of safe-using of this manual.
Following rules must be totally enforced to avoid accidences
The “Manual” is one of the elements of device components of “The company.
low-speed BenchtopAir-refrigerate Centrifuge”, which should be consistent with the
low-speed refrigerated centrifuges together in order to be accessed by the operator.
The The company. low-speed BenchtopAir-refrigerate Centrifuge is designed
for clinical medicine, biology, chemistry, genetic engineering, immunology and so on.
The maximum separation speed can not exceed the density of the sample 1.2g/cm³;
when the sample is greater than the density of 1.2g/cm ³, the maximum rotor speed
must be reduced accordingly.
When the The company. low-speed BenchtopAir-refrigerate Centrifuge process
is running (in the separation of the sample or when the rotor rotation), to make sure
that the operator must not stay within 30 cm around the centrifuge, no hazardous
substances or items blocking vents within 30 cm around the centrifuge.
If in the use of The company. low-speed BenchtopAir-refrigerate Centrifuge, the