English
User's Manual
of GNF001
Version 1, July 2023
- 15 -
When closing the door, please slowly close it with both hands and do not use
excessive force to prevent the door cover from hitting the palm or damaging the
locking hook!
5.5 Installing the rotor
The rotor used must be a product confirmed and designated by our company,
otherwise it may cause malfunction. This manual includes various rotor specifications
and models confirmed by our company. The company recommends users to use the
specifications and models of various rotors specified in this manual (see "Table 1:
Rotor Types and Technical Parameters" for details).
The use of inappropriate rotor types and centrifuge tube specifications
and other optional components can lead to poor centrifuge
performance and even damage to the centrifuge!
The steps for installing the rotor are as follows (as shown in Figures 5 and 6)
●
Turn on the power switch and wait for the self check to complete;
●
Press the door opening button
, open the centrifuge door cover, and
confirm that the cavity is clean, clean, and free of foreign objects;
●
Clean the surface of the motor spindle;
●
As shown in Figure 5, prepare the rotor you want to use. Hold the rotor
with both hands, align the center hole of the rotor with the motor
spindle, place it vertically, and place it at the bottom of the conical
surface. Release both hands, and then press the rotor down with your
hands.
●
Use the tool for disassembling the rotor (special Hex key) to tighten the lock nut
clockwise;