Press ENTER/PRINT
(1.4)
to save the value.
Press MODE/MEAS
(1.3)
to exit the constant adjustment mode and return to the mea-
suring mode.
Note: To cancel the constant adjustment process press MODE/MEAS (1.3)
- Cuando seleccione el modo Salinidad en pantalla visualizará "SALT"
4.50
SALT
%
25.0 ºC
Nota: para asegurar unos resultados precisos, es necesario antes de realizar las medi-
ciones, ajustar y establecer una serie de parámetros (ver Ajuste de parámetros).
Ajuste de parámetros
El conductímetro Nahita 910/5 permite ajustar diversos parámetros para adaptarse a
todo tipo de aplicaciones y asegurar unos resultados precisos:
Constante de la célula de conductividad (para determinaciones de conductividad, TDS
y salinidad)
Factor de conversión TDS (para determinaciones de TDS)
Coeficiente de temperatura (para determinaciones de conductividad y TDS)
1. Ajuste de la constante de la célula de conductividad
Cada célula de conductividad tiene su propio valor de constante; este valor deberá ser
introducido en el equipo cada vez que se cambie de célula. Para ello:
En cualquier modo de medida (Conductividad, TDS o Salinidad), presione
MODE/MEAS
(1.3)
Con los botones /C/T/S
(1.5)
o /SAVE
(1.6)
, seleccione en pantalla "CONT".
Presione ENTER/PRINT (1.4), de esta manera el equipo entra en el modo de ajuste de la
constante de la célula de conductividad.
Presione /C/T/S
(1.5)
para seleccionar el nivel en el que se encuentra el valor de la
constante de la célula que va a ser utilizada (0.01, 0.1, 1.0, 5.0 ó 10.0).
ENGLISH
CASTELLANO
Pág. 8
Manual de instrucciones 50910050 Revisión 1 Junio-07
- Cuando seleccione el modo TDS en pantalla visualizará "TDS"
98.5
TDS
mg / L
22.5 ºC
S
S
S
Version 1 June-07
Instruction manual 50910050 Page 25
2. TDS conversion factor adjustment
Nahita conductivity meter 910/5 takes by default a value of 0.50 for the TDS conversion
factor. The user can adjust this value within a range from 0.40 to 1.00. For this purpose,
proceed as follows:
In the TDS measuring mode press MODE/MEAS (1.3).
With /C/T/S (1.5) or /SAVE (1.6) select "TCOE" on display.
Press ENTER/PRINT (1.4), so as the equipment enters the TDS conversion factor adjus-
tment mode.
Press /C/T/S (1.5) (increase) or /SAVE (1.6) (decrease) to adjust the desired exact
value of the TDS conversion factor.
Press ENTER/PRINT (1.4) to save the value.
Press MODE/MEAS (1.3) to exit the conversion factor adjustment mode and return to
the measuring mode.
Note: TDS conversion factor can only be adjusted from the TDS measuring mode.
3. Temperature coefficient adjustment
Conductivity determination depends on temperature. The temperature variations can
be compensated by introducing a compensation factor (temperature coefficient) that will
be taken into account by the equipment to calculate the conductivity at the reference tem-
perature.
By default, the equipment takes a value of 0.020 (2 %/ºC) for the temperature coefficient;
if this needs to be modified, proceed as follows:
In the Conductivity or TDS measuring mode press MODE/MEAS (1.3).
With /C/T/S (1.5) or /SAVE (1.6) select "COEF" on display.
Press ENTER/PRINT (1.4), so as the equipment enters the temperature coefficient adjus-
tment mode.
Press /C/T/S (1.5) (increase) or /SAVE (1.6) (decrease) to adjust the desired exact
value of the temperature coefficient.
Press ENTER/PRINT (1.4) to save the value.
Pres MODE/MEAS (1.3) to exit the temperature coefficient adjustment mode and return
to the measuring mode.
Note: Temperature coefficient can only be adjusted from the Conductivity and TDS
measuring mode. The temperature coefficient for salinity determinations is already fixed
in the equipment and can not be modified.