2) The pump does not make a good vacuum or spends more time than expected.
CAUSE
SOLUTION
The oil is contaminated
Drain the contaminated oil and add new
oil
There are leaks or losses in the
Review the vacuum hose to detect
vacuum system
possible leaks
The level of oil is not enough
Stop the pump and add oil until reaching
the correct level if the oil is in perfect con-
ditions of use. If not, change the oil.
Loss of oil due to a defect of
Contact the Nahita Technical
the pump
Assistance Department
Internal parts of the pump damaged
Contact the Nahita Technical
or worn out by the use or a bad
Assistance Department
lubrication
3) The pump starts but presents a high electric consumption (amperage).
CAUSE
SOLUTION
The oil is too dense (high viscosity)
Use a lower viscosity oil
or environmental temperature is
lower than 5 ºC
The pump is overloaded with oil
Remove the excess of oil
Foreign bodies inside the pump;
Contact the Nahita Technical
broken vanes; damaged or blocked
Assistance Department
bearings
Low tension
Verify the tension used
4) The pump is very noisy when operating
CAUSE
SOLUTION
Connection worn out or loosen
Contact the Nahita Technical Assistance
Department
Bearings damaged
Contact the Nahita Technical Assistance
Department
Version 1 July-08
Instruction manual 5900110x
Page 13
the accessories supplied.
Please check that all the accessories are enclosed with the equipment:
- Oil bottle, 250 mL
- User’s manual
- Warranty certificate
We will only accept any equipment return within 15 days after delivery and provided
it comes in its original wrapping.
Installation
Before using the vacuum pump, it is convenient for you to familiarize with its compo-
nents and basic essentials, as well as with its control functions.
PLEASE READ THOROUGHLY THE INSTRUCTIONS BEFORE CONNECTING
AND OPERATING WITH THIS EQUIPMENT WITH THE AIM OF ACHIEVING
THE HIGHEST FEATURES AND THE MAXIMUM DURABILITY OF THE INS-
TRUMENT
Put the pump on top of a horizontal, plane and stable surface so as vibrations produced
during operation do not disturb nearby equipments. The back side of the pump must be
perfectly ventilated and free of obstructions.
Do not put the pump near any warm supply (burners, blowlamps…), nor expose it
directly to the sun, etc. During operation, flammable liquids must be out of the working
area.
Open the oil inlet cap
(1.5)
and fill the pump until the oil level reaches the middle of the
sight glass
(1.4)
.
Make sure that the switch
(1.2)
is off and insert the Schuko wire that feeds the AC elec-
tric current in the base of current 220 V, 50 Hz ± 10% provided with earth wire.
Neither the manufacturer nor the distributor will assume any responsibility for the
damages produced to the equipment during its installation or damages to persons suf-
fered by the improper use of the electric connection. The tension should be 220 V, 50
Hz ± 10%.
Setting up
High vacuum pumps are designed to offer highly satisfactory results for a long time
and don’t need a special assembly.
The aeration hose must be flexible to avoid a possible transmission of vibrations, water-
tight and with the same or higher calibre than the pump connection (depending on the
distance to cover) to obtain an optimum output.
Note: model 59001102 presents double vacuum connection. During operation, the con-
nection that is not in use must be perfectly protected by its cap (1.8) to avoid leaks.