27
ALLUX
TM
User’s Guide
Issue
Check item
Solution
Abnormal noise or
weakness in joints.
Check for foreign substances.
Are any screws etc. interfering with the
knee joint section?
Please remove any foreign matter and
ensure that normal movement takes
place.
Does vibration sound when connected
to the battery charger?
Please contact your prosthetist.
Any issue not outlined above
Please contact your prosthetist.
The safety lock
cannnot be applied.
Check that you holds ALLUX
TM
in a
slightly flexed state and apply the
weight sufficiently to ALLUX
TM
. (When
the weight is not applied or ALLUX
TM
is flexing or extending even slowly, the
lock does not function.)
Hold ALLUX
TM
in a sufficiently loaded
state.
In cases other than above
Please contact your prosthetist.
ALLUX
TM
is locked and
jams when you sit down.
Check that you do not stop flexing
ALLUX
TM
halfway.
Since the safety lock is effective, it
is necessary to sit down without
stopping.
Check that load is not applied to the
toe to flex ALLUX
TM
after transition to
the swing phase.
Since the stumbling lock is effective,
it is necessary to sit down without
applying load or sit down with the
yielding function.
In cases other than above
Please contact your prosthetist.
Troubleshooting
9
9
Troubleshooting
Содержание ALLUX NE-Z41
Страница 43: ...MEMO ...