19
ALLUX
TM
User’s Guide
Vibration alert
Alert conditions
Meaning
2
sec
When the battery charger or
Power OFF cap is connected
Control of ALLUX™ has been stopped.
0.25
sec ×
8
times
When the battery charger or
Power OFF cap are disconnected
The battery level supports at least 3 hours
0.5
sec ×
8
times
When the battery charger or
Power OFF cap are disconnected
The battery level supports less
than 3 hours
0.75
sec
8
times+
(4
sec
)
When the battery charger or
Power OFF cap are disconnected
Normal walking is not currently supported due to
battery level
*1
being insufficient. Please connect an
auxiliary battery or recharge it.
2
sec
Operational mode changed
via Remote Control App (See 5.3)
ALLUX™ mode change has been completed using the
Remote Control App.
0.5
sec *
3
times ×
15
times
Low remaining battery
High temperature state
Recovering from abnormally high
temperature When the control of
the hydraulic cylinder becomes
temporarily unstable
Status requiring a safe posture.
If this status continues ensure to contact your prosthetist
as a malfunction may have occurred.
4
sec *
2
times with
30
sec interval
When the main ALLUX™
unit is becoming abnormally hot
The main ALLUX™ unit is becoming abnormally hot.
However, normal walking can still continue.
4
sec *
4
times+
10
sec
When the main ALLUX™
unit has become abnormally hot
Normal walking is not currently supported by ALLUX™ and
will not return until the temperature has dropped. Please
wait until normal status has been achieved, as indicated
by vibration pattern ⑥.
0.25
sec *
50
times+
5
sec
When the main ALLUX™
unit has been malfunctioned
The status of ALLUX™ is abnormal and normal walking is
not supported. The abnormal status could be resolved by
inserting/removing the battery charger or power off cap.
If this does not solve the problem then please ensure to
contact your prosthetist.
2
sec
When the load is intermittently
applied to the toe or heel four
times
The wireless function of
ALLUXTM has been reactivated.
0.5
sec ×
1
times
Safety lock
The safety lock is being activated.
0.25
sec ×
8
times
When you press the restart button
of the Adjustment App or the
Remote Control App
Reboot ALLUX™.
*1: Remaining battery refers to the remaining capacity of ALLUX
TM
main battery or the backup battery pac. Only a guide as it can
vary with the usage conditions.
*2: If no vibratory notification occurs when the battery charger or power supply off connector is removed then an abnormality
may have occurred.
*3: If another notification condition is met while the alert is set to vibration mode, the vibration may be switched on.